Section 7. Software Reference
Baud Rate
Maximum Output Rate (Samples per Second)
19200
5
38400
10
57600
15
115200
20
Under
Options
, you can choose whether or not to output labels with each data
record, and whether to output diagnostic text records. An example of a data record
sent with and without labels is shown below.
Data format with labels:
(Data (Ndx 87665) (DiagVal 757) (Date 2009-09-10) (Time 14:06:44:140) CO2Raw
0.0332911) (H2ORaw 0.19299) (CO2D 5.20672) (H2OD 755.566) (Temp 15.517)
(Pres 99.4361)…
Data format without labels:
87665 757 2009-09-10 14:06:44:140 0.0332911 0.19299 5.20672…
The End of Line Delimiters determine the character(s) that terminate the data
records. Depending on your recording device, a line feed, carriage return, or both
may be required to properly parse the data records.
Under
Data Output
, select the data records that you want to output; click
Select All
to choose all, or
Select None
to disable all checked values.
Ethernet Output
The
Ethernet
tab is used to set the instrument's Ethernet port network com-
munication. After configuration, click
Apply
and the analyzer will begin to send
data out the Ethernet port according to these parameters.
Use the
Update Rate
menu to select from .1, .2, .5, 0, 1, 2, 5, 10, or 20 Hz.
Under
Options
, you can choose whether or not to output labels with each data
record, and whether to output diagnostic text records. An example of a data record
sent with and without labels is shown below.
Data format with labels:
7-14
LI-7200RS Enclosed CO₂/H₂O Analyzer
Summary of Contents for LI7200RS
Page 1: ...Enclosed CO2 H2O Gas Analyzer Instruction Manual...
Page 2: ......
Page 11: ...ix Appendix C Suppliers Appendix D Configuration Grammar Appendix E Warranty Appendix F Index...
Page 12: ...x...
Page 20: ...Section 1 General Information 1 8 LI 7200RS Enclosed CO H O Analyzer...
Page 54: ...Section 3 Operation 3 12 LI 7200RS Enclosed CO H O Analyzer...
Page 112: ...Section 4 User Calibration 4 12 LI 7200RS Enclosed CO H O Analyzer...
Page 130: ...Section 5 Maintenance 5 18 LI 7200RS Enclosed CO H O Analyzer...
Page 138: ...Section 6 Troubleshooting 6 8 LI 7200RS Enclosed CO H O Analyzer...
Page 184: ...Section 8 Theory of Operation 8 12 LI 7200RS Enclosed CO H O Analyzer...
Page 187: ...Gas Analyzer Dimensions Appendix A Specifications A 3 Appendix A Specifications...
Page 188: ...Appendix A Specifications LI 7550 Dimensions A 4 LI 7200RS Enclosed CO H O Analyzer...
Page 192: ...Appendix B Pin Assignments B 2 LI 7200RS Enclosed CO H O Analyzer...
Page 196: ...Appendix C Suppliers C 4 LI 7200RS Enclosed CO H O Analyzer...
Page 225: ......