Setting the span calibration
Follow suitable spanning procedures, as described in
Plumb a span gas to the analyzer air inlet. Use a span gas balanced in air (do not use
a nitrogen-balanced gas).
1
To start a span calibration, enter:
mosquitto_pub -h 172.24.82.42 -t "licor/niobrara/system/CO2_Span"
-m "<CO2 tank in ppm>"
l
mosquitto_pub
launches the application to publish (send to the instrument).
l
-h
specifies a host (the Trace Gas Analyzer) followed by the instrument IP
address
l
172.24.82.4x
is the IP address. You can also enter the host name. Both are
given on the instrument display.
l
-t
specifies that you want to subscribe to a topic.
l
"licor/niobrara/system/CO2_Span"
is the topic and hierarchy for the LI-
7815.
l
-m
specifies that you are sending a message from the command line.
l
"<CO2 tank in ppm>"
is the CO
2
concentration in PPM.
This initiates the span calibration in the LI-7815. The span is implemented imme-
diately.
2
To check if the span is complete, enter the command:
mosquitto_sub -h 172.24.82.42 -t "licor/niobrara/cfg/CO2_Span"
The instrument will reply with the span value (near 1) like
"1.00242"
3
Type
Ctrl + C
to stop the script.
Section 7. Communication with the MQTT protocol
7-5
Setting the span calibration
Summary of Contents for LI-7815
Page 1: ...LI 7815 CO2 H2O Trace Gas Analyzer Instruction Manual...
Page 2: ......
Page 10: ...viii...
Page 30: ...Section 2 Getting started 2 14 LI 7815 CO H O Gas Analyzer...
Page 38: ...Section 3 Applications of the LI 7815 3 8 LI 7815 CO H O Gas Analyzer...
Page 42: ...Section 4 Software overview 4 4 LI 7815 CO H O Gas Analyzer...
Page 54: ...Section 4 Software overview 4 16 LI 7815 CO H O Gas Analyzer...
Page 62: ...Section 5 Troubleshooting 5 8 LI 7815 CO H O Gas Analyzer...
Page 70: ...Section 6 Maintenance 6 8 LI 7815 CO H O Gas Analyzer...
Page 88: ...Appendix B Additional safety information B 6 LI 7815 CO H O Gas Analyzer...
Page 92: ...Appendix C Specifications C 4 LI 7815 CO H O Gas Analyzer...
Page 97: ......