Section 4. Software overview
l
IP Address:
The assigned IP address is displayed here (and on the instrument dis-
play).
Static:
Used to set a static IP address. This setting is used when configuring the
instrument to communicate over a cellular gateway, such as the Sierra Wireless RV-
50X AirLink. When selected, you manage the networking manually by setting the IP
address, subnet mask, and possibly the gateway.
l
IP Address:
Enter the desired IP address. For a local network, you will use some-
thing like 172.24.1.nn, where the values represented by nn are octets that
uniquely identify this instrument on the network.
l
Subnet Mask:
Typically 255.255.255.0.
l
Default Gateway:
Set the first two octets to match those of the IP address.
Wi-Fi
Under
Settings > Wi-Fi
you can configure the wireless network settings, including
whether the wireless network is on or off, and which channel it uses. You can always
use the default setting, but if there are many devices using that channel, you may
have better results with a channel that has fewer devices. If you have difficulty con-
necting, try changing the channel that the instrument uses.
4-8
LI-7815 CO₂/H₂O Gas Analyzer
Summary of Contents for LI-7815
Page 1: ...LI 7815 CO2 H2O Trace Gas Analyzer Instruction Manual...
Page 2: ......
Page 10: ...viii...
Page 30: ...Section 2 Getting started 2 14 LI 7815 CO H O Gas Analyzer...
Page 38: ...Section 3 Applications of the LI 7815 3 8 LI 7815 CO H O Gas Analyzer...
Page 42: ...Section 4 Software overview 4 4 LI 7815 CO H O Gas Analyzer...
Page 54: ...Section 4 Software overview 4 16 LI 7815 CO H O Gas Analyzer...
Page 62: ...Section 5 Troubleshooting 5 8 LI 7815 CO H O Gas Analyzer...
Page 70: ...Section 6 Maintenance 6 8 LI 7815 CO H O Gas Analyzer...
Page 88: ...Appendix B Additional safety information B 6 LI 7815 CO H O Gas Analyzer...
Page 92: ...Appendix C Specifications C 4 LI 7815 CO H O Gas Analyzer...
Page 97: ......