Calibration Issues
CO2 and H2O Analyzers
Using the LI-6400
18-13
Setting the H
2
O Span
To check the span of the H
2
O analyzer, youÕll need a known concentration of
H
2
O. The best choice for this is the LI-COR Dew Point Generator (LI-610).
The H
2
O IRGA gain adjustment is a process by which the user can manually
(using the arrow keys
-¯
) adjust the values of
G
wr
and
G
ws
(see Equations
(14-5) and (14-6) on page 14-5).
■
■
■
■
To set the H
2
O Span
1
Setup the LI-610 for an appropriate dew point
Subtract about 5C from room temperature, and use that for the target dew
point temperature. Wait until the condensorÕs temperature (as monitored on
the LI-610) reaches this target.
The reason for this 5C ÒbufferÓ is to avoid condensation in the line between
the LI-610Õs condensor and the IRGA. If condensation happens, you will
have large errors.
2
Set the flow rate.
Use about a 0.5 l min
-1
flow from the LI-610.
3
Attach to the IRGA
You have the same choice, as shown in Figure 18-6 on page 18-12. Here,
however, we would recommend option B: splitting the flow and connecting
both the reference and sample, with Match Off. The reason for this is that you
will be able to drastically reduce the equilibrium time, waiting for the sample
cell, as you will see.
4
View water channels
Press
F3
or
F4
to make the water IRGAs the active ones.
5
Wait for equilibrium
If you are using Option A (connected to sample, Match On), then be prepared
to wait about 20 minutes, until the rise in sample and reference concentration
is negligible.
If you are plumbed for Option B (sample and reference connected), ignore the
If you do not have an LI-610, or some device of similar
accuracy, do
not
adjust the IRGA span values for water.
Summary of Contents for LI-6400
Page 1: ...Using the LI 6400 Portable Photosynthesis System ...
Page 15: ...Part I The Basics ...
Page 16: ......
Page 174: ...Making Measurements Answers to Questions 4 56 Using the LI 6400 4 ...
Page 175: ...Part II Useful Details ...
Page 176: ......
Page 200: ...Standard Tools Power ON Hooks 5 24 Using the LI 6400 5 ...
Page 214: ...Real Time Data Real Time Graphics 6 14 Using the LI 6400 6 ...
Page 234: ...Environmental Control Light Control 7 20 Using the LI 6400 7 ...
Page 244: ...Light Sensor Considerations Gallium Arsenide Phosphide GaAsP Sensor 8 10 Using the LI 6400 8 ...
Page 288: ...Data Logging Making Your Own AutoPrograms 9 44 Using the LI 6400 9 ...
Page 289: ...Part III Working With Files ...
Page 290: ......
Page 312: ...The LPL File System Troubleshooting 10 22 Using the LI 6400 10 ...
Page 340: ...Downloading Files Using a Data Capture Program 11 28 Using the LI 6400 11 ...
Page 375: ...Part IV Configuration Issues ...
Page 376: ......
Page 420: ...Defining User Variables Old Style vs New Style 15 18 Using the LI 6400 15 ...
Page 454: ...Using an Energy Balance Further Reading 17 12 Using the LI 6400 17 ...
Page 455: ...Part V Maintenance Troubleshooting ...
Page 456: ......
Page 572: ...Troubleshooting Useful Information 20 46 Using the LI 6400 20 ...
Page 593: ...Part VI Programming ...
Page 594: ......
Page 622: ...Programming with LPL Compiler Directives 22 28 Using the LI 6400 22 ...
Page 846: ...Index I 16 Using the LI 6400 ...