User switch adjustments
The user switch-button can be pressed (as seen below) by using the adjusting-pin that is added in the transmitter
package or by using another pin with similar shape (Ø 1mm).
The color of the LED-ring/LED will indicate the status of the transmitter during user switch adjustments:
Transmitter status
LED-ring/LED color
Normal operation
Solid green
Vacuum-zero/Full-scale adjustment
Pulsing between red and green
If the user switch is activated by accident and vacuum Zero or Full scale adjustment is executed the
original factory adjustment can be recovered using the FD!VAC or FD!ATM command. Please refer
to the Communication Protocol 300544663 (RS232) or Communication Protocol 300544664 (for
EtherCAT) for further details.