5.5
Operation Via Serial Interface
5.5.1
Description
Transmission parameters
Type
RS232 / RS485
Transmission rate
9600 baud
Data bits
8
Stop bits
1
Parity bit
0 (none)
Communication mode
Half-duplex
Bus operation
2 devices (recipient, transmitter)
The transmission parameters of the control system (PC, PLC) must agree
with the above transmission parameters.
Waiting time between sending and receiving a string must be ≥50 ms.
Communication
Only ASCII characters can be transmitted. Each data transmission is terminated with
<CR><LF>. For data transmission to the CMOVE 1250, no blanks (SPACE) are admitted.
All characters have to be upper case. Since there is no time limit between the individual
signs, manual operation via the interface is possible.