www.levtech.ro

Temperature transmitter
for Pt100 or Pt1000

LSP-TSTS

Summary of Contents for LSP-TSTS

Page 1: ...User Manual Lueta Nr 319 Romania 40758576007 o ce levtech ro www levtech ro Temperature transmitter for Pt100 or Pt1000 LSP TSTS...

Page 2: ...g 6 2 5 Cleaning 7 3 DESCRIPTION 8 4 FEATURES 8 5 INSTALLATIONAND WIRING 9 5 1 Installation onTS 35 DIN Rail 9 5 2 Connections with ScrewTerminals 9 6 SPECIFICATIONS 10 7 CONFIGURATIONTYPE AND ORDER C...

Page 3: ...erns about the operation and maintenance of this product please contact our customer support If a problem arise with the product check if you followed the installation steps correctly If the problem p...

Page 4: ...pprovals Date Version Number Document Changes 11 03 2021 1 0 Initial draft 13 04 2021 2 0 Overall improvements 15 04 2021 3 0 Application examples added Name Title Levi Mihaly General Manager Document...

Page 5: ...onal injury may result from not knowing and following these instructions In order to ensure long product life Levtech recommends that you utilize the product and perform maintenance by correctly follo...

Page 6: ...n the manual and who are able to follow these should connect the device Mounting and connection of the device should comply with national legislation for mounting of electric materials i e wire cross...

Page 7: ...xplosive atmospheres disconnect power before servicing and do not separate connectors when energized and an explosive gas mixture is present Do not mount or remove devices from the power rail when an...

Page 8: ...temperature sensor connected on its input On the input it is possible to connect both 3 wires and 2 wires Pt100 or Pt1000 sensors This transmitter is suitable for any industrial application and with...

Page 9: ...ttach the module in the lower part of the rail as shown in the left gure below 2 Press the module upwards 1 Apply leverage using a screwdriver as shown in the right gure below 2 Press the module downw...

Page 10: ...t Signal type Current Output Range current type 4 20mA on current loop Output Resistanc 75 Output current noice 30nAPP Output accuracy 0 1 Power Supply Supply Voltage 16 30VDC Power Consumption 40mA R...

Page 11: ...ad 24V In 4 20mA Loop 2 3 1 5 6 4 4 20mA Loop Pt100 wire 1 Pt100 wire 2 Pt100 wire 3 2 3 1 3 wire PT100 2 3 1 2 wire PT100 Jumper 16 30V In 77 5 mm 48 3 mm 6 5 mm 100 00 mm 8 9 TERMINAL DESIGNATION DI...

Page 12: ...ample In the below example a Pt100 temperature sensor is connected to our transmitter and the output signal is connected to a Siemens Logo analog module on 4 20mA current loop 10 ELECTRICAL DIAGRAMS 1...

Page 13: ...e that you have a Pt100 temperature sensor and you have to measure the temperature in a water tank and control the heater of the water tank with a PLC You can transmit the measured value to the PLC us...

Page 14: ...defect will be made free of charge If you purchased the item from our distributors and it is under warranty please contact them to notify us of the situation Levtech Service Department will determine...

Page 15: ...vtech COMPANY DATA levtech_s_p levtechsp 14 CONTACT INFORMATION ADDRESS Lueta Nr 319 Post Code 537140 Harghita Romania LEVTECH SERVICE PRODUCTION SRL Fiscal code CIF RO35733217 Trade Register RC J19 1...

Reviews: