User Manual for LEVIFLOW
®
LFSC-D Flowmeters
www.levitronix.com
PL-4507-00, Rev08, DCO# 20-024
18
3.2.3 Instructions for Mechanical Installation
!
WARNING
Hazardous voltage may be present.
In order to avoiding fluid spills shorting voltages within the converter,
place it in a spill protected environment (for example protected
electronic cabinets).
If explosive flammable gases are present, place the converter in an
explosion-proof cabinet.
1. The converter shall be installed in a spill protected cabinet.
2. Assure that sufficient ventilation is provided in order to avoid exceeding the allowed ambient
temperature and humidity range specified in
3. Bring the fixation bars into un-lock position (see
4. If multiple converters are used (see
) stack them together.
5. Attach the connector box (see
6. Attach the termination box (
) and fix it by pushing the fixation bars of the termination
) to the inside.
7. Bring the fixation bars into the locked position (see
8.
Mount the whole stack on a DIN rail following the same steps as described in
the single channel converter
LFC-1C-PC
.