www. levelprosales.com
The
ITC 450/470
meter is equipped with one current input (0-20mA), two voltage inputs (0-10V and 0-150
mV), one RTD input (Pt 100/500/1000) and one TC input (thermocouple: K, S, J, T, N, R, B, E). Temperature of
cold ends is compensated automatically. RTD and TC inputs are fully linearised. It is allowed to use only one
input at the same time. Input ranges are described in the next chapter. Result is showed on 4-digit LED
display.
The device can be equipped with two or four relay (or OC type) outputs. Optionally
ITC 450/470
with two
relays (or OC type) outputs can be equipped with active current output, passive isolated current output or
active voltage output. Device
ITC 450/470
is equipped with RS-485 / Modbus RTU communication interface
and sensor supply output. The meter can be ordered in two power supply versions.
ITC 450/470
can be used for controlling and regulation of processes need proportional and threshold control
like: temperature processes (heating or cooling), valves controlling or other.
Power supply voltage
(depending on version)
85...
230
...260V AC/DC; 50 ÷ 60 Hz (separated)
or 19...
24
...50V DC and 16...
24
...35V AC (separated)
External fuse (required)
T - type, max. 2 A
Power consumption
max. 6.5 VA @ 85 ÷ 260V AC/DC
max. 6.5 VA @ 16V ÷ 35V AC
max. 6.5 W @ 19V ÷ 50V DC
Current input (20 mA)
0÷20 mA, 4÷20 mA overload protected, input current is limited to 50 mA (typically)
Current measurement accuracy
± 0.1% @ 25°C; ± one digit (for 0÷20 mA range)
Current input resistance
< 65 (typical 30 )
Accepted prolonged
input overload
20%
Voltage input (10V range)
0÷5 V, 1÷5 V, 0÷10 V, 2÷10 V
Voltage measurement accuracy
± 0.1% @ 25°C; ± one digit (for 0÷10 V range)
Voltage input resistance
> 100 kW (while maintaining correct polarization)
Accepted prolonged
input overload
20%
Voltage input (150mV range)
0÷60 mV, 0÷75 mV, 0÷100 mV, 0÷150 mV
Voltage measurement accuracy
± 0.1% @ 25°C; ± one digit (for 0÷150 mV range)
Voltage input resistance
> 1,5 M
Accepted prolonged
input overload
20%
3. TECHNICAL DATA
2. GENERAL CHARACTERISTICS
02
www. levelprosales.com
ITC 450/470
User manual
Summary of Contents for ITC 450
Page 14: ...www levelprosales com 12 www levelprosales com ITC 450 470 User manual...
Page 17: ...www levelprosales com 15 www levelprosales com ITC 450 470 User manual...
Page 38: ...36 www levelprosales com ITC 450 470 User manual 7 4 MENU STRUCTURE...
Page 39: ...37 www levelprosales com ITC 450 470 User manual...