7.7.2 Enabling ST Analysis/Display ST Parameters
The ST monitoring function is disabled by default. Before you start
ST monitoring, enable the ST analysis function. To do so, follow the
steps below:
1. Enter the ST setup screen in either of the following ways:
Select the ST parameter area.
Press the [Parameters] quick key, select the [ST] tab.
Press [Main Menu]
→
[Parameters], select the [ST] tab.
2. Switch on [ST Analysis], the ST parameter area appears on the
main screen.
The ST parameter display area is configured differently according to
the ECG cable used:
When you are using the 3-lead ECG leadwires, a separate ST
parameter area does not appear on the display. The ST
deviation value displays in the ECG parameter area.
When you are using the 5-lead ECG leadwires, the ST
parameter area displays 7 ST deviation values for lead I, II, III,
aVR, aVL, aVF, V.
When you are using the 6-lead ECG leadwires, the ST
parameter area displays 8 ST deviation values for lead I, II, III,
aVR, aVL, aVF, Va, Vb.
When you are using the 12-lead ECG leadwires, the ST
parameter area displays 12 ST deviation values for lead I, II, III,
aVR, aVL, aVF, V1, V2, V3, V4, V5, V6.
This example shows the ST parameter area with 5-lead ECG. Your
monitor screen may look slightly different from the illustration.
1
Parameter label
2
ST measurement unit
3
ST alarm off symbol
4
Lead labels
1
4
2
3
5
Summary of Contents for AIView 12V
Page 1: ...AIView Series Patient Monitor Operator s Manual...
Page 2: ......
Page 6: ...IV This page intentionally left blank...
Page 24: ...1 7 This page intentionally left blank...
Page 26: ...2 2...
Page 34: ...2 10 2 6 5 Bottom View 1 2 2 2 2 1 Battery compartment 2 Anti slip foot pad...
Page 35: ...2 11 This page intentionally left blank...
Page 44: ...3 9 1 2 3 4 5 8 9 7 6...
Page 55: ...3 20 This page intentionally left blank...
Page 71: ...5 11 This page intentionally left blank...
Page 78: ...6 7 This page intentionally left blank...
Page 106: ...8 5 This page intentionally left blank...
Page 112: ...9 6 This page intentionally left blank...
Page 119: ...10 7 This page intentionally left blank...
Page 129: ...11 10 2 Set PR Source select an appropriate PR source from the drop down list...
Page 165: ...15 12 Other sources of interference if any...
Page 166: ...15 13 This page intentionally left blank...
Page 194: ...17 21 This page intentionally left blank...
Page 204: ...19 5 This page intentionally left blank...
Page 210: ...20 6 This page intentionally left blank...
Page 229: ...23 10 This page intentionally left blank...
Page 248: ...25 8 This page intentionally left blank...
Page 251: ...A 3 Shutdown Delay At least 5 minutes since the first low battery alarm...
Page 283: ...D 2 This page intentionally left blank...
Page 289: ...E 6 less than 3V m...