A-5
2 TEMP connectors
2 IBP connectors
1 CO
2
connector
Thermal
recorder
Thermal dot matrix recorder
Horizontal resolution: 8 dot/mm
Vertical resolution: 8 dot/mm
Recording paper width: 50 mm
±
1 mm
Paper speed: 25 mm/s, 50 mm/s, error
≤±
5%
A.5 Signal Outputs Specifications
Auxiliary output
Standard
Meets the requirements of IEC 60601-1 for
short-circuit protection and leakage current
ECG analog output
Bandwidth
(-3dB; reference
frequency: 10Hz)
0.5 Hz ~ 40 Hz
QRS delay
< 25ms(non-paced)
Output sensitivity
0.4 V/mV5%
Baseline level
2.20
±
0.2 V
Output impedance
≤
100
Ω
Duration of
short-circuit
Short circuit to ground for 1 minute, no fault after
disconnection.
Defibrillation synchronization signal output
Output impedance
≤
100
Ω
Maximum time
delay
30 ms
Amplitude
High level: 3.5 V to 5 V, providing a maximum of 10
mA output current;
Low level: < 0.5 V, receiving a maximum of 5 mA
input current.
Pulse width
100 ms
±
10%
Summary of Contents for AIView 12V
Page 1: ...AIView Series Patient Monitor Operator s Manual...
Page 2: ......
Page 6: ...IV This page intentionally left blank...
Page 24: ...1 7 This page intentionally left blank...
Page 26: ...2 2...
Page 34: ...2 10 2 6 5 Bottom View 1 2 2 2 2 1 Battery compartment 2 Anti slip foot pad...
Page 35: ...2 11 This page intentionally left blank...
Page 44: ...3 9 1 2 3 4 5 8 9 7 6...
Page 55: ...3 20 This page intentionally left blank...
Page 71: ...5 11 This page intentionally left blank...
Page 78: ...6 7 This page intentionally left blank...
Page 106: ...8 5 This page intentionally left blank...
Page 112: ...9 6 This page intentionally left blank...
Page 119: ...10 7 This page intentionally left blank...
Page 129: ...11 10 2 Set PR Source select an appropriate PR source from the drop down list...
Page 165: ...15 12 Other sources of interference if any...
Page 166: ...15 13 This page intentionally left blank...
Page 194: ...17 21 This page intentionally left blank...
Page 204: ...19 5 This page intentionally left blank...
Page 210: ...20 6 This page intentionally left blank...
Page 229: ...23 10 This page intentionally left blank...
Page 248: ...25 8 This page intentionally left blank...
Page 251: ...A 3 Shutdown Delay At least 5 minutes since the first low battery alarm...
Page 283: ...D 2 This page intentionally left blank...
Page 289: ...E 6 less than 3V m...