8-3
Pre-sampling: acquire the ECG data of 10 seconds before
pressing the [Start] button. At the completion of acquisition, the
system enters the preview screen.
In the preview screen, press [Save] to send the saved data to
AI-ECG server for analysis. After the report is obtained, you can view
it in the [Reports Mgmt] screen.
In addition to the above 12-leadECG analysis, the monitor also
supports 12-lead analysis of stored historical waveforms. Follow this
procedure:
1. Press [Review] quick key or press [Main Menu]
→
[Review] to
enter the review screen.
2. Select the [Full Disclosure] tab.
3. Select the desired waveform segment.
4. Press the [12-Lead ECG] button to enter the resting ECG
acquisition screen.
5. Set the [Period].
6. Press [Start].
The monitor will send the stored data corresponding to the selected
time length, or the collected data of the selected time length, to the
AI-ECG server (if connected) for analysis.
Note
[Start Time] is the time point of the selected waveform segment.
8.5 Reports Management
After ECG acquisition, press [Main Menu]
→
[Reports] to enter the
reports management screen. You can perform the following
operations in this screen:
Press [Delete] to delete one or more selected patient record.
Press [Preview] to enter the preview screen of ECG waveforms
of the selected patient record.
Press [Upload] to upload one or more selected patient record to
the AI-ECG server.
Press [Download] to download the diagnosis report of the
selected patient record from the AI-ECG server.
Press [View Report] to view the diagnosis report of the selected
patient record.
Summary of Contents for AIView 12V
Page 1: ...AIView Series Patient Monitor Operator s Manual...
Page 2: ......
Page 6: ...IV This page intentionally left blank...
Page 24: ...1 7 This page intentionally left blank...
Page 26: ...2 2...
Page 34: ...2 10 2 6 5 Bottom View 1 2 2 2 2 1 Battery compartment 2 Anti slip foot pad...
Page 35: ...2 11 This page intentionally left blank...
Page 44: ...3 9 1 2 3 4 5 8 9 7 6...
Page 55: ...3 20 This page intentionally left blank...
Page 71: ...5 11 This page intentionally left blank...
Page 78: ...6 7 This page intentionally left blank...
Page 106: ...8 5 This page intentionally left blank...
Page 112: ...9 6 This page intentionally left blank...
Page 119: ...10 7 This page intentionally left blank...
Page 129: ...11 10 2 Set PR Source select an appropriate PR source from the drop down list...
Page 165: ...15 12 Other sources of interference if any...
Page 166: ...15 13 This page intentionally left blank...
Page 194: ...17 21 This page intentionally left blank...
Page 204: ...19 5 This page intentionally left blank...
Page 210: ...20 6 This page intentionally left blank...
Page 229: ...23 10 This page intentionally left blank...
Page 248: ...25 8 This page intentionally left blank...
Page 251: ...A 3 Shutdown Delay At least 5 minutes since the first low battery alarm...
Page 283: ...D 2 This page intentionally left blank...
Page 289: ...E 6 less than 3V m...