Password protection and basic program
21
Password protection and basic
program
A four-digit password is required to enter data and to carry out
settings in the basic program. The password is the current time of
Testomat 2000
®
Polymer
in reverse order.
Password entry
Press the "M" key
The basic menu "
" appears
Use the cursor block to select the desired menu item
"
"
The selection appears in CAPITAL LETTERS.
Press "ENTER" to confirm your selection
The selected submenu "
" appears
Press "ENTER" to confirm the menu item "
"
The cursor flashes in the
Use the cursor keys to enter the time in
reverse order
:
Press "ENTER" to confirm the entry
The selection menu for the basic program appears. You can now
enter the plant specific data.
Selecting the operating mode
Under the menu item
it is possible to select
the type of analysis controller. Testomat 2000
®
Polymer
provides
numerous selection options: Time control, quantity control via water
meter and external analysis start.
Shortest interval = 0 minutes between analyses. Largest interval =
255 minutes.
The analysis interval (time between two analyses) is determined by
the duration of the supplementary program AUX, the set flush times
(internal and external), the programmed interval and the duration of
the analysis.
Selecting the time control
In
, select
=>
=>
=>
Press "ENTER" to confirm the selection
(An asterisk " * " appears at the end of the line)
("TIME CONTROLLED" " * " is preset)
▼▲
▼▲
▼▲
Please enter the
time in reverse
order:
=> 5201
Testomat instrument
(display) in display mode
Time controlled
Internal start via a
timer.
▼▲