98
TS11/TS15, Check & Adjust
Laser guide screws
Laser Guide adjust-
ment step-by-step
This step-by-step description describes the Laser Guide adjustment for a distance of
50 m. Place the target plate at a distance of 120 m to perform the laser guide adjust-
ment for 120 m.
a) Horizontal adjustment screw
b) Fixing screw
c)
Fixing screw
d) Horizontal adjustment screw
e) Vertical adjustment screw
f)
Vertical adjustment screw
g) Safety cover screw
h) Safety cover
a b
c d
e
f
e
f
g
h
TS_124
Step
Description
)
Make sure that the instrument is levelled.
1.
Place the laser guide target plate at a distance of 50 m and aim with the
telescope of the instrument at the crosshairs of the target plate.
2.
Loosen the safety cover screw (g) and move the safety cover (h) to the
side to access the vertical adjustment screws.