250
Troubleshooting
•
ATTO655 for 640 nm
•
Rhodamine 6G for 532 nm or fluorescein (in a solution with a high pH value)
•
ATTO488 for 470-nm excitation
• The molecular brightness is the count rate of the detectors divided by the number of
molecules present in the detection volume. It is displayed in the FCS preview. In general,
a 10-nM solution is used. For a day-to-day comparison, use the same laser power, ideally
controlled using a power meter. A more detailed description of FCS measurements is
provided in the point measurement section.
31.2.1
Causes for Decreased Performance
If a performance drop is noticed, the origin of the performance drop needs to be determined
first. A drop in system performance may be caused by:
• Decreasing laser power:
Check the laser intensity with a laser power meter. If the laser power from PQ lasers is
too low, either the laser coupling unit has to be readjusted or the laser needs to be
repaired or replaced. In this case, please contact customer support.
• Decreasing detection sensitivity:
•
Misalignment of the TCS SP8: Contact Leica Service.
•
Misalignment of the fiber decoupler for the fluorescent light: Contact Leica Service.
•
Misalignment of the detector(s): In this case, please contact Leica customer support.
31.3
No FLIM Image is Displayed During Measurement
Can laser light be seen over the objective during measurement?
If not:
• Is the shutter open in LAS AF?
• Has the laser been activated by moving the corresponding slider?
• Is the multifunction port in the correct position?
• Is the pulse repetition rate of the pulsed lasers correct? This can be checked in the
Time
Trace
preview window. When starting a measurement, at least background noise should
be displayed (see
at the right in
). If the repetition rate is equal to
0, check whether the SYNCHRONIZATION is set to external on the PDL800B or PDL800D
and whether the laser emission lamp is illuminated during scanning.
If yes:
• Do you see a dark count rate when activating the
Time Trace
tab in the
Preview
window? If not, check to ensure that the detectors are still activated. The DSN 102 (if
present) deactivates a detector automatically if the count rate exceeds a certain limit. In
Observe additional manuals
For more information, refer to the manual for the PDL800B or D.
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......