![Leica TCS SP8 SMD User Manual Download Page 224](http://html1.mh-extra.com/html/leica/tcs-sp8-smd/tcs-sp8-smd_user-manual_1873063224.webp)
224
Changing Detector Connections on the Scan Head and Router When Using HyD-RLD
24.2
Connect and Use Detectors
24.2.1
Using SP FLIM PMT
If you would like to use SP FLIM PMT, you have to connect the SMD signal cable of the
detector in the scan head with the router. Proceed as follows:
1. Close the LAS AF software.
2. Switch the system, including the detectors, off.
3. Connect the SMD signal cable of the detector in the scan head (
) to the
).
4. Connect the synchronization cable of the MP or white light laser with Channel 0 of the
PicoHarp 300 counting unit (
). The connection between Channel 1 and the
PHR 800 router can remain.
5. Switch the system, including the detectors, on.
6. Start the LAS AF software and select the hardware tree
MP on FCS FLIM 2 APD
or
MP
off FCS FLIM 2 APD
.
24.2.2
Using MPD APDs
If you would like to use MPD APDs, you have to connect the image signal cable and the SMD
signal cable of the detector with the scan head and router. Proceed as follows:
1. Close the LAS AF software.
2. Switch the system, including the detectors, off.
3. Connect the image signal cable of the MPD APDs (SPAD 1 and SPAD 2,
) to the
signal converter (
). The cables can always stay attached to the signal
converter.
4. Connect the signal converter to the APD printed circuit board of the scan head
5. Connect the SMD signal cable MPD APDs (
) to the PHR 800 router's detector
connections (
). These cables can always stay attached to the router.
6. Connect the synchronization cable of the MP or white light laser with Channel 0 of the
PicoHarp 300 counting unit (
). The connection between Channel 1 and the
PHR 800 router can remain.
7. Switch the system, including the detectors, on.
8. Start the LAS AF software and select the hardware tree
MP on FCS FLIM 2 APD
or
MP
off FCS FLIM 2 APD
.
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......