S Series User Manual
Quick Start Guide 36
If you set the diopters on the adjustable
eyepiece exactly as described, the
image will remain equally sharp and constant
(parfocal) from the lowest to the highest
magnification. This means you do not have
to refocus when changing magnification. The
focus needs to be readjusted only if you want to
view a specimen detail that is located higher or
lower. Use this advantage as often as possible, it
is not available on all stereo microscopes.
The diopters can be set b5
and -5.
The following adjustments have to be
carried out only once by each user. Using
reticules leads to slightly deviated settings,
which are described in the user manuals on the
reticules (measurement).
Adjusting the diopter settings
1. With Leica S APO, turn the rotary knob to
the "Vis" setting.
2. Turn the dioptric correction on the
adjustable eyepiece to the center position.
Continued on next page
Diopter Settings and Parfocality: 1 Adjustable & 1 Fixed Eyepiece
Summary of Contents for S APO
Page 1: ...Leica S Series User Manual...
Page 10: ...S Series User Manual Leica S Series 10 Leica S Series...
Page 16: ...S Series User Manual Assembly 16 Assembly...
Page 26: ...S Series User Manual Quick Start Guide 26 Quick Start Guide...
Page 45: ...S Series User Manual Get Set 45 Get Set...
Page 52: ...S Series User Manual Extend 52 Extend...
Page 70: ...S Series User Manual Photography Video 70 Photography Video...
Page 73: ...S Series User Manual Dimensional Drawings in mm 73 Dimensional Drawings in mm...
Page 78: ...S Series User Manual Technical Data 78 Technical Data...