Instrument Setup
4
4.
Instrument Setup
4.1
Site requirements
Warning
The instrument must not be operated in areas at risk of explosion.
To ensure proper function of the instrument, it must be set up while maintaining a minimum distance
of 10 cm from walls and furniture.
•
The instrument requires an installation area of approx. 650 x 550 mm.
•
Relative humidity 20 –
85
% maximum - non-condensing
•
Room temperature consistently b15 °C and +30 °C
•
Elevation: up to max. 2000 m NN
•
The instrument is designed for indoor use only.
•
The power plug/circuit breaker must be freely and easily accessible.
•
Power supply at a distance no greater than the length of the power cord – an extension cable must
not be attached.
•
The bench must have a sufficient load capacity and rigidity with respect to the weight of the
instrument.
•
Avoid vibrations, direct sunlight, and large temperature fluctuations. The installation location must be
well ventilated, and must contain no sources of ignition of any kind.
•
The instrument must be connected to a grounded power socket.
•
Only
the power cord cord that is provided with the instrument and that
is intended for the local power
supply may be used.
•
The installation location must be protected against electrostatic discharges.
4.2
Unpacking the instrument
Warning
When the device arrives, check the tilt indicator
of the arrow is blue, the shipment was not handled as prescribed.
In this case, please mark the shipping documents accordingly and check the shipment for damage!