Adjusting the Circular Level of the Instrument and Tribrach
18447_001
1
2
3 - 4
1.
Place and secure the instrument into the tribrach and onto a tripod.
2.
Turn on the instrument. Using the tribrach footscrews, level the
instrument with the electronic level.
☞
Electronic level:
•
Select
Devices
from the Home Menu.
•
Tap the arrow button to the right of the device name.
•
To display the electronic level, tap
Compensator
.
3.
Check the position of the circular level on the instrument and tri-
brach.
☞
If both circular levels are centred, no adjustments are necessary.
4.
If one or both circular levels are not centred, adjust as follows:
Instrument
: If it extends beyond the circle, use the supplied allen
key to centre it with the adjustment screws. Turn the instrument by
200 gon (180°). Repeat the adjustment procedure if the circular
level does not stay centred.
Tribrach
: If it extends beyond the circle, use the supplied allen key
to centre it with the adjustment screws.
☞
After the adjustments, all adjusting screws must have the same
tightening tension and no adjusting screw should be loose.
Adjusting the circular
level step-by-step
40
Check & Adjust