13
Leica CV5030 - Robotic Coverslipper
4.1
Location requirements
The equipment location must meet the following requirements:
•
Stable, plane, vibration-free lab table.
•
Largely vibration-free floor.
•
Space of approx. 25 cm on the left side of the unit (to open the side
door).
•
Space behind the unit for connecting the fume tubing.
•
Room temperature generally between +10 °C and +35 °C.
•
Relative humidity 80% maximum, non-condensing.
The use of the unit in hazardous areas is not allowed.
4.2
Setup and initial operation
4.
Initial setup
Fig. 3
Perform the repacking of the unit in reverse
order to the unpacking instructions. The pack-
ing material should, therefore, always be kept
(unpacking instructions on the outside of the
transport packaging).
Mountant bottle must be taken from the unit.
Remove all residues from the system using a
suitable solvent.
The unit must be transported in upright posi-
tion only as indicated by labels on the outside
of the transport box.
To lift the unit, pick it up by reaching under the
sides (at least 2 persons are required—the
device weighs approx. 57 kg) and lift it onto a
solid laboratory table.
•
Pull the protective plastic cover upwards to
remove it.
•
Ensure that the delivered accessories are
complete in accordance with the order.
The chemicals used in the Leica
CV5030 are easily inflammable and
hazardous to health. The installation
location of the CV5030 must be well-
ventilated; there should be no igni-
tion sources there of any kind.
Adjust the height of the threaded feet (Fig. 3) to
ensure that the unit is level in all directions.
Summary of Contents for CV5030
Page 2: ......