Annex
ARveo / Ref. 10 747 384 / Version 04 77
16 Annex
16.1 Checklist before the operation
Patient
.......................................................................................................................................................................................
Surgeon
.......................................................................................................................................................................................
Date
.......................................................................................................................................................................................
Step Procedure
Details
Checked / Signature
1
Cleaning the optical accessories
X
Check the tubes, eyepieces and the documentation
accessories (if used) for cleanliness.
X
Remove dust and dirt.
2
Installing the accessories
X
Position the handles as desired.
X
Connect the mouth switch and/or footswitch if used.
X
Check the camera image on the monitor and realign if
necessary.
3
Checking the tube settings
X
Check the tube and eyepiece setting for the selected user.
4
Balancing
X
Balance the ARveo (see page 25).
X
Press the "All Brakes" button on the handle and check the
balancing.
5
Function check
X
Check the fiber optics cable connection to the optics
carrier.
X
Connect the power cable.
X
Switch on the microscope.
X
Switch on the illuminator at the control unit.
X
Leave the illumination on for at least 5 minutes.
X
Check the lamp history and make sure that the remaining
life time is sufficient for the planned surgery.
X
Replace defective bulbs before the surgery.
X
Test all functions on the handles and the footswitch.
X
Check the user settings on the control unit for the selected
user.
6
Safety check
X
Check counterweights and accessories for secure seating.
7
Positioning at the OP table
X
Position the ARveo on the OP table as required and lock
the foot brake (see page 31).
8
Sterility
X
Fit sterile components and sterile drape if used
(see page 32).
9
Final work
X
Check that all equipment is in its proper position
(all covers fitted, doors closed).