Operation
34
ARveo / Ref. 10 747 384 / Version 04
8 Operation
8.1 Switching the microscope on
WARNING
Danger of fatal electrical shock.
X
The ARveo surgical microscope may be connected to a
grounded socket only.
X
Operate the system only with all equipment in its proper
position (all covers fitted, doors closed).
WARNING
Danger of injury to the eyes due to possibly hazardous
optical infrared and UV radiation.
X
Do not look at the operating lamp.
X
Minimize exposure to eyes or skin.
X
Use appropriate shielding.
WARNING
Danger of burn injuries in otologic surgery.
X
Use the lowest comfortable light intensity.
X
Adjust the field of view to match the operating field.
X
Irrigate the wound frequently.
X
Cover the exposed parts of the pinna with a moist
surgical sponge.
X
Connect the microscope to a grounded socket.
X
Do not position the microscope to make it difficult to operate
the disconnection device, which is the mains plug.
X
Switch on the microscope at the power switch (2) on the stand.
After the surgical microscope is switched on, the settings of the
last active user are loaded.
X
Check the fiber optics cable connection to the optics carrier.
X
Switch on the illumination with the key (1) on the control unit.
1
2
The main screen is displayed.
3
X
Check both lamp hour counters by switching from lamp 1 to
lamp 2 with button (3).
To guarantee a good light performance the life time shall not
exceed 500 hours.