![Leica BIOSYSTEMS HistoCore BIOCUT Instructions For Use Manual Download Page 22](http://html.mh-extra.com/html/leica-biosystems/histocore-biocut/histocore-biocut_instructions-for-use-manual_1873468022.webp)
22
Version 1.7, Revision K
Instrument Components and Specifications
3
Rear view
Fig. 5
1
2
3
1
Magnet
3
Retraction switch
2
Coarse feed direction change
3.2
Instrument specifications
Basic instrument with 2-step mechanical trimming, silent retraction with ON/OFF selection,
ergonomically positioned coarse feed wheel with user selectable turn directions, without specimen
orientation or additional accessories.
• Manual mechanical rotary microtome with low-maintenance and backlash-free precision micrometer
feed system.
• Horizontal feed and vertical stroke mechanisms with cross roller bearings.
• Smooth-running hand-wheel allows two manual sectioning modes: rocking mode and conventional
manual sectioning with full hand-wheel rotation.
• Two independent hand-wheel locking systems.
• Operator-adjustable force balancing system with spring force compensation offers two advantages:
1. Flexibility to adapt the spring force to different weights of specimen/clamps, minimizing the risk
of an object head dropping into the knife.
2. No need for a heavy counter balance in the hand-wheel.
• Silent specimen retraction system with ON/OFF function.
• Ergonomically positioned coarse feed wheel which can be personalized by user selectable preferred
turn direction.
• Ergonomically positioned 2-step mechanical trimming lever for quick trimming of blocks using
predefined trim steps of 30µm and 10µm.
• Large top surface area allows placement of objects that require a flat surface.
• Removable top tray allows storage of sectioning tools and prevents items from falling.
3.3
Technical Data
General
Section thickness range
1 - 60 µm
Summary of Contents for HistoCore BIOCUT