14. Status Display
14-30
14.6.4
Setting Gamut Error Detection Levels
To set gamut error detection levels, press F•1 GAMUT in the status menu.
The settings that you make here also apply to the R, G, and B bars in the 5 bar display.
Reference: Section
12.6.1, “5 Bar Display Explanation”
STATUS
→
F•5 ERROR CONFIG
→
F•4 ERROR LEVEL
→
F•1 GAMUT
→
Figure 14-19 GAMUT menu
• Setting the Upper Limit
To set the gamut error upper limit, follow the procedure below. An error is detected when
the SDI input signal level exceeds the specified value.
If you press F•D, the gamut error upper limit will be reset to its default value of 109.4.
Procedure
STATUS
→
F•5 ERROR CONFIG
→
F•4 ERROR LEVEL
→
F•1 GAMUT
→
F•1 GAMUT
UPPER %
Settings
Selectable range: 90.8 to 109.4 (The default setting is 109.4.)
• Setting the Lower Limit
To set the gamut error lower limit, follow the procedure below. An error is detected when
the SDI input signal level goes below the specified value.
If you press F•D, the lower limit will be reset to its default value of –7.2 %.
Procedure
STATUS
→
F•5 ERROR CONFIG
→
F•4 ERROR LEVEL
→
F•1 GAMUT
→
F•2 GAMUT
LOWER %
Settings
Selectable range: –7.2 to 6.1 (The default setting is –7.2.)