- 38 -
12. MAIN BOARD JUMPER SETTINGS
Serial Signal
Duplex
Jumpers
Termination Resistor*
RS485
Full
None
E6 a = Transmit
E6 c = Receive
Half
E6 b + d**
E6 c
RS232
Full
None
None
*
The termination resistor jumper settings should only be selected if the transmitter is the last
device on an RS485 line longer than 200 feet (60 m).
** Or connect external BTX to BRX and ATX to ARX (same effect as internal jumpers).
To reset communications
to 9600 baud, command mode, Custom ASCII protocol, and Address 1,
place a jumper at E1 and power up the transmitter.
Analog Output
J4 Pins
Jumpers
Current, 4-20 mA
1 Lo, 2 Hi
E2 a + d
Voltage, 0-10V
1 Lo, 2 Hi
E2 b + c
Voltage, -10V to +10V
3 Lo, 2 Hi
E2 b + c
Jumper settings are for CTR main board Rev J
Excitation Output*
Jumpers
5V, 100 mA
E3 a + c; E4 a
10V, 120 mA
E3 a + c; E4 b
24V, 50 mA
E3 b, E4 none
*
Attempting to draw more than the rated excitation output current will shut down the output.