ECO Gold
67 / 134
10
Technical data and graphs
The figures were determined according to DIN 12876.
Data applicable to all ECO GOLD thermostats
Ambient temperature range
°C
5 – 40
Relative humidity
Maximum relative humidity 80 % at 31 °C and decreasing linearly to 50 %
up to 40 °C.
Contamination level
2
Setting resolution
K
±0.01
Display resolution
K
±0.01
Accuracy of indication
K
±0.2
Pump type/number of power levels
Pressure pump/6
Discharge pressure, max.
bar
0.55
Discharge flow, max.
L/min
22
Viscosity of the heat carrier liquid
mm²/s
Heating range: maximum 150;
Control range:
≤
30
Display field
TFT display 3.5"; 320 x 240 pixel
Programmer
5 programs with a total of 150 temperature/time segments
Standard interface
USB
Class of protection
IP 21
Class designation
lII
Marking
FL
(
suitable for flammable and non-flammable liquids)
Overvoltages
Overvoltage Category II and transient overvoltages according to Category II.
Class of protection for electrical operat-
ing equipment DIN EN 61140 (VDE
0140-1)
Class I
Summary of Contents for Viscotemp 15 G
Page 77: ...11 Declaration of Conformity 09 2021 ECO Gold 77 134 ...
Page 78: ...78 134 ECO Gold 09 2021 ...
Page 81: ...09 2021 ECO Gold 81 134 W Warnings 53 92 X Xp 106 Xpf 108 ...
Page 82: ...82 134 ECO Gold 09 2021 ...
Page 83: ...09 2021 ECO Gold 83 134 ...
Page 84: ...84 134 ECO Gold 09 2021 ...
Page 85: ...09 2021 ECO Gold 85 134 Appendix with settings ...
Page 121: ...09 2021 ECO Gold 121 134 Click on the tab with the mouse and then on the Device manager ...
Page 129: ...09 2021 ECO Gold 129 134 ...
Page 130: ...130 134 ECO Gold 09 2021 ...
Page 131: ......