Proline Low-temperature Thermostats
60
Starting up
YACE0072 / 21.08.07
−
Select the desired resolution with
or
.
−
Accept selection with
or End ,
or
quit the window with
without making
changes.
7.8 Special
settings
7.8.1 Setpoint
resolution
This function enables the resolution of the setpoint T
set
to be increased from the standard value of
0.1 °C to 0.01 °C.
Master
R
3x
to
LMDls
to
O@p@
1x
−
Call setpoint resolution
R
.
−
The current setting is displayed (here
<0
for the works setting of
0.1°C).
−
The resolution display
<0
flashes.
or
−
<0
for 0.1°C
−
</0
for 0.01°C
Wait 4 seconds or
−
Display flashes 4 s
Æ
new value is automatically accepted,
or
−
the value is immediately accepted with the Enter key.
7.8.2 Defining the type of start mode
Usually it is desirable that the thermostat carries on operating again after an interruption in the voltage
supply. However, if for safety reasons you do not wish this, you can insert an intervening manual
activation step.
R <0
°C
Help Menu End T
set
T
fix
0,1
0,01
0,001
Resolution
Summary of Contents for RP 845
Page 2: ......