Proline Kryomats
37 / 137
7
Starting up
Warning!
Addition of liquids with low boiling points (e.g. water to hot oil),
alteration of liquid properties
(reducing the flash point)
Explosion, burns, scalds, fire
Site the device in suitable premises.
Avoid dripping water and condensation.
Do not position any small parts and liquids above the device.
Keep the cover on the thermostat (if present) closed.
Prevent the ingress of secondary liquids (e.g. from
customer's heat exchanger).
Do not work with liquids in the direct vicinity of the device.
Check the heat transfer liquid at least every six months (e.g.
mixing ratio with a hydrometer).
Warning!
Risk of heat transfer liquid overheating due to incorrect entry of
overtemperature switch-off point T
max
.
Fire
The overtemperature switch-off point must be below the
flash point of the heat transfer liquid.
Set the overtemperature switch-off point (T
max
) to 5 K
above the upper limit of the temperature range for your
application.
Warning!
Bursting of the external consumer due to overpressure (e.g. glass
apparatus)
Explosion
Install the tubes so that they do not kink.
For consumers with a maximum permissible operating
pressure below the maximum pressure of the pump, use a
pressure relief device for protection.
Warning!
Risk of refrigerant circuit bursting from excessive ambient
temperatures while device is inoperative
Explosion
Observe the permitted storage and operating temperatures.