7.11.1
Internal control variable (integral measurement probe)
Only read further here, if you have no external temperature probe connected (and activated according to Section
7.5.4 as control variable).
Command
Control Parameters
Open the device parameter menu via the soft
key Menu .
With the cursor keys continue to
Control
Control Parameters
Control Parameters
.
The adjacent window appears.
Change parameters marked with (auto) where
necessary to manual input with
Tv manual/auto
.
Select the parameters to be changed with
and confirm with
.
Then in the following settings window, change
the value and confirm with
.
7.11.1.1
Proven settings for control parameters and pump (integral measurement probe)
Instrument Type Heat transfer liquid
Xp
_P
Tn
tn
Tv
tu
Td
td
Pump level
RP 4090 CW
Water
9.0
60
42
6.3
8
RP 4090 CW
Ethanol
9.0
60
42
6.3
8
RP 4090 C
Water
9.0
60
42
6.3
8
RP 4090 C
Ethanol
9.0
60
42
6.3
8
Technical changes reserved!
7.11.2
External control variable (External measurement probe)
You only need to read further here if you have connected an external temperature probe or the actual temperature is
read in from a module (and you have activated it as control variable according to Section 7.5.4).
Only modify the control parameters if you have knowledge of control techniques.
The control system for external actual values is implemented for improvement of the control behavior as a two-stage
cascade controller. A "master controller" determines the "internal setpoint", from the temperature setpoint and the
external temperature, passed to the slave controller. The control value of the slave controller controls the heating and
cooling.
Pump Menu
End
T
set
T
fix
Control Parameters
Control para. sets
Tv manual/auto
Self Adaption
Correction limitation
Xp
6,0
Tn
30
Tv
(auto)
21
Td (auto)
3,5