112 / 137
Proline Kryomats
9.2.2
Draining the water-cooled condenser
Important: With the risk of frost (e.g. transport in winter), drain the condenser on water-cooled
devices.
Remove the water hose on the water tap. Than open the solenoid valve for the water as described below. Blow
compressed air in the water return hose Continue until all water has flowed out of the device.
Command
Start unfill condenser
Open the device parameter menu via the soft
key Menu .
Change from
Pump
Start unfill conden.
using .
Use
to open the solenoid valve for the
water. The display changes automatically to the
overview window.
To close the valve again selects this menu again.
The item is called now
End unfill conden.
.
9.3
Servicing, cleaning, repair and disposal information
Warning!
Live parts during troubleshooting
Electric shock hazard
Disconnect the device from the mains before the repair (e.g.
when changing components).
Have the repair carried out only by a specialist.
Warning!
Rotating / live parts when removing the ventilator fan
Cuts, crushing, electric shock hazard
Disconnect the device from the mains before the repair.
Have the repair carried out only by a specialist.
Settings
Graph
Clock
Programmer
Interfaces
Control
Limits
Pump Menu
End
T
set
T
fix
Pump Level
Ext.Pump
Start unfill conden
Pump