110 / 137
Proline Kryomats
9
Maintenance
9.1
Device status
The thermostat can be conveniently checked with the Command remote control.
9.1.1
Interrogating the device type
Settings
Device status
Device type
9.1.2
Software version
Settings
Device status
Software version
.
The versions of the control system (
Control
), safety system (
Safety
), Command remote control (
Command
),
cooling system (
Cool
) and, where applicable, other connected modules are displayed.
9.1.3
Serial numbers
Settings
Device status
Serial numbers
.
The serial number of the Master (
Master
), Command remote control (
Command
), cooling system (
Cool
) and
other connected modules are displayed.
9.1.4
Device data
Command
Device data
Settings
Device status
Device
data
Display
Text shows various actual temperatures in °C
from ext. Pt100 and the modules.
T cont. head and T heatsink are temperatures of
electronics in the Master in °C.
Pump power in Watts, speed in rpm, current in
ampere (A).
T
int
indicates the current internal bath
temperature in °C.
Mains voltage in percentage (%) of nominal and
frequency in hertz (Hz).
Level indicates the liquid level in the internal
bath.
Voltage of power transformer, 5V supply and
fan voltage in Volt.
Cur. cons.: Mains current consumption in
Ampere.
T ext Pt
25.70 Tint
-8
T ext analog
---.-- Mains U(%) 100.74
T ext serial
---.-- Mains frequ.
50
T cont. head 39.80 Level
4
T heatsink
51.68 Low voltage 27.90
Pump pow.
44.90 5V supply
5.00
Pump rpm
5460 Fan voltage
7.0
Pumpe cur.
1.68 Cur. cons.
2.84
Pump Menu
End
T
set
T
fix