Proline Low-temperature Thermostats
YACE0072 / 21.08.07
Starting up
57
Command
Control Variable
−
Open the device parameter menu with the
soft key Menu .
−
With the cursor keys, change further to
Æ
Control
Æ
Control Variable
.
−
Internal
is currently active.
−
Select other control variables (only
displayed when present) with
or
and confirm with
or End ,
−
or quit the window with without making
any changes.
7.7.5 Current consumption from the mains
If your mains fuse is rated below 16 A, the current consumption can be reduced in steps from 16 A to
10 A using this function. Of course, the maximum heating power of 3.5 kW is then also reduced
accordingly. Take into account whether other loads are still connected to the fused circuit or whether
your Proline Thermostat is the only load.
or
Change temperature probe source:
−
0
for the internal probe,
−
DO
only when an external probe is connected,
−
D@
only when an analog module is connected and configured,
−
DR
only when a serial module is connected and is continuously
receiving actual values from a PC.
Wait 4 seconds
or
−
Display flashes 4 s
Æ
new value is automatically accepted,
or
−
Value is immediately accepted with Enter key.
−
If
DO
D@
or
DR
has been selected,
then the green
LED indicates that the control has regulated to
the external temperature signal.
Help Menu
End T
set
T
fix
Internal
External Pt100
Analog module
Digital module
Control Variable
01!34
°C