![Lauda RE 1050 G Operating Instructions Manual Download Page 86](http://html1.mh-extra.com/html/lauda/re-1050-g/re-1050-g_operating-instructions-manual_4088318086.webp)
ECO GOLD
86
Maintenance
28.04.10/ YACE0088
9.3.4
Displaying and changing the device type
The device type (
Ì
4.2) without the suffix "G" (GOLD) is shown in the menu under
Setup
Æ
Device Status
Æ
Type
.
A type change can be carried out here in this menu point.
The type change is linked to a 3 second key depression delay to prevent an unintentional type change. In
this way all parameters are reinitialised, i.e. even adjusted control parameters are lost.
Automatic adaptation of Tmax to the selected device type, i.e. with the ECO GOLD thermostat Tmax =
202 °C and with the ECO GOLD with transparent bath Tmax = 102 °C.
Following this, Tmax must be re-entered manually (
Ì
7.4.1), because otherwise the device goes into the
error status (error message in ECO GOLD "T max diff. Ctrl-Safety".
9.3.5
Displaying serial numbers
The corresponding serial numbers of Control and Safety are displayed in two lines under the menu point
Serial numbers. Provided they are available, the serial numbers of connected modules are also dis-
played.
−
Access to the menu level is obtained by pressing
.
−
Selection and confirmation of
Æ
Setup
Æ
Device Status
Æ
Serial numbers
.
ESC
סּ
OK
Standby
Serial numbers
Control
LCK1913-09-0002
Cool
KTExy-08-9876
Ext Pt
Summary of Contents for RE 1050 G