ECO GOLD
28.04.10/ YACE0088
Maintenance
83
kk
−
If the bath temperature increases above the overtempera-
ture switch-off point:
Two-tone alarm signal sounds.
"Overtemperature" appears in the display.
Heating switches off on both poles.
Pump and chiller are switched off via the electronics.
−
Rectify the cause of the malfunction.
−
Wait until the bath temperature has cooled below the cut-
off point or set the cut-off point higher than the bath tem-
perature. If "Overtemperature" appears in the display:
−
Unlock
with
.
−
With longer periods of unsupervised operation you should
check the overtemperature
protection; to do this:
−
Slowly reduce T
max
as described in (
Ì
7.5.1). The thermostat should switch off when the actual tem-
perature is greater than T
max
.
−
An alarm message (step 1-2, see above) should follow.
−
Reset the switch-off point to be higher than the bath temperature.
−
Unlock
with
.
9.1.2
Low level protection alarm and low level protection check
kk
When the liquid level falls so far that the heaters are no longer
completely covered with liquid, an alarm is triggered.
Two-tone alarm signal sounds.
"Low Level Pump" appears in the display.
Heating switches off on both poles.
Pump and chiller are switched off via the electronics.
−
Rectify the cause of the malfunction and top up the missing
liquid (
Ì
−
Unlock
with
.
−
Check the safety system at regular intervals
by lowering the bath level.
−
An alarm message (step 1-2, see above) should follow.
−
Top up with heat transfer liquid.
−
Unlock
with
.
No. Alarm
1
Low Level Pump
1
Low Level Pump
Display
סּ
OK
Standby
No.
Alarm
3
Overtemperature
3 Overtemperature
Display
סּ
OK
Standby