Proline Edition X cooling thermostats
82
YACE0100 / October 4, 2018
7.14.2 Low-level alarm and low-level checking
If the liquid level falls so far that the heating element is no longer
completely covered with liquid, an alarm is initiated:
1. The alarm sounds as a dual-tone signal.
2. Display for
LEUEL
(low level) is shown when the bath
contains too little liquid.
3. The red LED
above the fault triangle
flashes.
Heater switches off on both poles;
Varioflex pump and refrigerating unit are switched off.
Find the cause of the fault and, where necessary, top up for
missing liquid (
Press the Enter key.
Also, press this key if the unit has been switched off in the fault
state.
Checking the safety system at regular intervals
by lowering the bath level. To do
this, push hose onto pump connector and pump heat transfer liquid into a suitable
vessel.
Step 1
– 2 must follow.
With this test, the bath temperature must not be below 0 °C or above max. 50 °C,
otherwise there is a risk of burning!
If irregularities arise during the checking of the safety devices, switch off the unit
immediately and pull out the mains plug.
Have the equipment checked by LAUDA Service Constant Temperature Equipment.
Command
Low-level alarm!
Low-level alarm is shown in the display and signifies that
unlocking is only possible on the Master control panel.
Level alarm
LEUEL
°C