Proline Edition X cooling thermostats
YACE0100 / October 4, 2018
53
7.8.7 Entering the offset of the internal temperature probe
If, during checking with a calibrated reference thermometer probe a deviation is found, then the offset
(i.e. the additive part of the characteristic) of the internal measuring chain can be adjusted with the
following function. The reference thermometer must be dipped into the bath according to the details on
the calibration certificate.
Master
CAL 1
3x
to
MNEnu
2x
to
CAL..
Call the offset adjustment of the internal temperature probe
CAL..
.
CAL 1
is displayed.
3 seconds
The actual value of the bath temperature is displayed flashing.
or
Set the value which you have read from the calibrated reference
measurement device (with glass thermometers consider the
correction where applicable).
The entered value is accepted and
donE
is displayed.
Command
Calibration
Open the device parameter menu via the
soft key Menu .
With the cursor keys continue to
Settings
Calibration
intern Pt100
Calibration
.
The window shown on the left appears.
Confirm selection with
.
CAL 1
°C
Pump Menu
End
T
set
T
fix
intern Pt100
extern Pt100
Calibration
Default
Summary of Contents for Proline RP 1290 C
Page 2: ......
Page 4: ......
Page 115: ...Proline Edition X cooling thermostats YACE0100 October 4 2018 115 Blank page ...
Page 116: ...Proline Edition X cooling thermostats 116 YACE0100 October 4 2018 Blank page ...
Page 118: ......
Page 119: ......