Proline heating thermostat
57 / 123
7.8.10
Restoring the works setting of the external temperature-probe offset
If the offset has been misadjusted unintentionally, the works setting can be restored with this function.
Master
dEF E
3x
to
MNEnu
2x
to
CAL..
3x
Call the offset works setting for the external temperature probe
dEF
E
.
Continue as described in (
7.8.8) for the internal temperature probe.
Command
Default
Open the device parameter menu via the soft
key Menu .
With the cursor keys continue to:
Settings
Calibration
extern Pt100
Default
.
The adjacent window appears.
Confirm selection with .
Continue as described in (
7.8.8) for the in-
ternal temperature probe.
Pump Menu
End
T
set
T
fix
intern Pt100
extern Pt100
Calibration
Default
Summary of Contents for PROLINE P 5 C
Page 13: ...02 2022 Proline heating thermostat 13 123 Clear view thermostat PVL 24 Bridge thermostat PB C ...
Page 113: ...02 2022 Proline heating thermostat 113 123 Pump characteristics measured with water ...
Page 114: ...114 123 Proline heating thermostat 02 2022 Pump characteristics measured with water ...
Page 115: ...02 2022 Proline heating thermostat 115 123 12 Declaration of Conformity ...
Page 118: ......
Page 119: ......