Maintenance
YAUE0007
/ 2/26/2009
9.1.5 Fault memory (Command)
For the analysis and localization of faults the Command version includes a fault memory in which up to
45 fault and alarm messages are saved.
Command
−
Error store
Æ
Settings
Æ
Device status
Æ
Error
store
Æ
Display
.
−
The last message is at the top.
−
Each message line can be marked with
the cursor keys. The message appears in
plain text in the footer.
−
Under “Source”, the CAN node is
displayed which signaled the fault.
−
Code is the number, which in the Master
is shown in the display until the cause
has been rectified.
−
Type: Alarm, Warning or Fault (Error).
9.2
Cleaning
9.2.1 Cleaning the surface of the device
Withdraw the equipment mains plug before cleaning!
Cleaning can be carried out with water to which a few drops of surfactant (washing-up liquid) have been
added and using a damp cloth.
No water must enter the control section!
)
Carry out appropriate decontamination if hazardous material is spilt on or in the
equipment.
The cleaning or decontamination method is determined by the user's specialist
knowledge. In case of doubt, contact the manufacturer.
No. Source Code Type Date
Time
10 Safety 2 Alarm ------------ -----------
9 Safety 4 Warn. 28.08.03 15:32:02
8 Contro. 32 Error 17.07.0310.:52:02
7 Contro. 3 Warn. 06.06.03 11:15:11
6 Contro. 9 Alarm 05.06.03 08:45:01
5 Contro. 3 Alarm 01.06.03 17:58:22
4 Contro. 4 Warn. 28.05.03 20:01:22
3 Contro. 5 Warn. 27.05.03 07:58:00
Low level
Help Menu
End
T
set
T
fix