This section is relevant for:
Device operation using the Command Touch unit
1.
Open the Home window.
2.
Touch the [Tih / Til] tile.
3.
Touch one of the following buttons:
Select the first entry
Til
to set the lower limit value.
Select the second entry
Tih
to set the upper limit value.
4.
Press [OK] to confirm.
Refer to
“Input window for entering values manually” on page 67 for more
information on working with the input windows.
5.10
Setting the temperature target value T
set
The temperature target value T
set
is the temperature that the constant
temperature equipment should reach and then maintain.
This section is relevant for:
Device operation using the Command Touch unit
1.
Open the Home window.
2.
Touch the [Tset] button.
The entry window appears. A temperature target value can be
entered within the limit values displayed.
3.
Enter the new temperature target value.
4.
Press [OK] to confirm the new value.
The value is accepted.
Chapter 5.6.2 “Entry window” on page 67
information on working with the input windows.
Setting Tih and Til
Fig. 38: Selecting the Temperature limit tile
Fig. 39: Entering target value
V6
PRO bath thermostats and circulation thermostats
74 / 156