Designation
Abbreviation
Unit
Hold-back time
Tve
s
Attenuation time
Tde
s
The following control parameters can be set on the slave controller (P-con-
troller).
Designation
Abbreviation
Unit
Proportional range
Xpf
K
If Tv manual/auto is set to auto, Tve, Tde and Prop_E cannot be
modified. In this case, they are derived with fixed factors of Tne.
Prop_E is a constant predefined value here.
The following parameters may also influence the external control:
Temperature limits: Til and Tih
Chapter 5.9 “Setting temperature
limits Tih and Til” on page 73
Controller output limit: Heating power and cooling capacity
6.5 “Defining the actuating signal limit” on page 89
“Correction limitation” on page 100
6.8.4 Opening the control menu
1.
Open the home window.
2.
Touch the
Settings
Control
tile.
6.8.5 Saving a parameter set
You have set the control variable to
Internal
or
External
. This depends on
whether you wish to select internal or external control parameters.
1.
Select the
Control parameters
menu item in the Control menu.
2.
Modify the control parameters in the menu. See
“Editing internal control parameters” on page 102 and
6.8.7 “Editing external control parameters” on page 102 for a corre-
sponding description.
3.
Select the
Save as...
menu item in the Control parameters menu.
Press the [Save as…] button to open the alphanumeric keyboard.
4.
Enter a name for the parameter set. You can choose any name and it
can contain up to 16 characters.
5.
Press the [OK] button to save this name automatically.
If the name has already been used, the old parameter set is
overwritten.
Control parameters on the slave controller
V6
PRO bath thermostats and circulation thermostats