4.2 External consumer
4.2.1 Hoses
CAUTION!
Discharge of heat transfer liquid during operation caused by use
of unsuitable hoses
Frostbite
Use hoses with temperature resistance that is appropriate
for the operating temperature range of the device.
CAUTION!
Contact with cold hoses
Frostbite
Use insulated hoses for temperatures below 0 °C.
The hoses specified below can be used for all heat transfer liquids
that are approved for the devices.
Further information about the pump connections of the individual
Chapter 11.4 ‘Hydraulic circuit’
Tab. 3: Hoses, not insulated
Type
Hose
nozzle
Maximum permissible
pressure
Clear width
Ø in mm
External
diameter in
mm
Tem-
perature
range in °C
Catalogue
number
EPDM hose
10 mm
Devices with maximum
pump pressure of < 1 bar
9
11
10 – 90
RKJ 111
EPDM hose
½
" (13 mm) Devices with maximum
pump pressure of < 1 bar
12
14
10 – 90
RKJ 112
Rubber hose with fabric
reinforcement
½
" (13 mm)
Up to 10 bar
13 (
½
")
19
-40 – 100
RKJ 031
Rubber hose with fabric
reinforcement
¾
" (19 mm)
Up to 10 bar
19 (
¾
")
27
-40 – 100 RKJ 032
Approved tubes, adapters and hose clips
V05REV08
Microcool
24 / 57