7 Operation
7.1
Selecting operating modes
The constant temperature equipment has two operating modes:
Operation
: The constant temperature equipment’s components are
used as intended.
Standby
: All functions of the constant temperature equipment are
deactivated.
This operating mode is suitable for adjusting multiple settings, for
example.
A program that has been started is paused in Standby mode. If the
program is to continue later, it must be continued manually after
switching to Operation mode.
1.
Open the Home window.
2.
To interrupt the current mode, tap the [Stop] button in the bottom
right for 1 second.
The constant temperature equipment is now in
Standby
mode.
The [Stop] button has been replaced by the [Play] button.
3.
To restart operation, tap the [Play] button in the bottom right for
1 second.
The constant temperature equipment starts to operate again. You
can now see the [Stop] button in the bottom right.
7.2 Programmer
7.2.1 Basic information
The programmer allows you to save a temperature-time program. A program
consists of several temperature-time segments. A segment contains spec-
ifications relating to program replays, temperature, duration, pump level,
control variable and behavior of the switching outputs. Ramps, temperature
jumps and temperature maintenance phases are possible.
You can start a program manually (
Chapter 7.2.2 “Using programs”
on page 32), or you can configure a timer and select the required program
Chapter 7.5.1 “Timer” on page 36. In total, you can save up to 100 tem-
perature-time programs. Each program can consist of up to 250 freely
programmable segments.
Switching operating modes
Fig. 21: Programs tile
V1
Command Touch operating unit
27 / 51