6.3.1 Limiting the current consumption
If your mains fuse is less than 16 A, the current consumption of the constant
temperature equipment can be reduced incrementally from 16 A to 8 A.
The maximum heating power is reduced accordingly. Establish whether other
consuming units or only your device is connected to the fused circuit.
1.
Open the Home window.
2.
Touch the
Settings
Basic settings
Current consumption
tile.
An input window for entering values manually appears.
3.
Change the current consumption accordingly.
4.
Press [OK] to confirm the new value.
The value is accepted.
6.3.2 Safe Mode
Safe Mode
is a safe temperature state to which the constant temperature
equipment can be switched as soon as any malfunctions or errors occur.
Temperatures considered to be safe depend on the type of application
connected and should be entered on the LAUDA constant temperature
equipment before using the device for the first time. In the event of an error,
switching to Safe Mode can be carried out manually or automatically.
When Safe Mode is active, the constant temperature equipment continues
to run but the values preset here for the T
set
set temperature and control
variables are automatically set. In the Safe Mode submenu, you specify
how the constant temperature equipment should react to certain events
(actions):
Which events lead to the activation of Safe Mode?
Being activated manually via the constant temperature equipment.
Command via interface.
Being triggered by certain alarms.
Communication with the control station being interrupted.
What indicates that Safe Mode is activated?
The
Safe Mode
tile displays a yellow symbol.
What happens in active Safe Mode if no alarm is pending?
The
Safe Mode
tile displays a yellow symbol.
Safe Mode is ended by tapping the tile. The tile displays a gray symbol.
What happens in active Safe Mode if no alarm is pending?
The Touch Display shows the alarm window. The
Safe Mode
tile displays
a yellow symbol.
Safe Mode can only be ended by tapping the tile when the alarm on the
constant temperature equipment has been unlocked.
Fig. 13: Current consumption 16 amperes at
present
V1
Command Touch operating unit