7.
Wait until the displayed bath temperature T
int
matches the set
temperature T
set
.
8.
Select the
Menu
Setup
Calibration
intern Pt1000
2-Punkt upper
menu item.
9.
Enter the temperature value read off the reference thermom-
eter into the device.
10.
Press [OK] to confirm the new value.
The upper value has been accepted. 2-point calibration
has been completed.
Use this menu item to restore the calibration configured at the fac-
tory.
1.
Select the
Menu
Setup
Calibration
intern Pt1000
Factory calibration
Yes
menu item.
This deletes the customer's calibration and restores the
calibration as it was configured in the factory.
5.8 Restore factory setting
Use this menu item to restore the factory settings in the device.
With
All default
the control parameters, internal Pt1000, and
other parameters are reset to factory settings.
With
Control parameter
only the control parameters are reset
to the factory setting.
With
internal Pt1000
only the calibration of the internal tem-
perature probe is reset to the factory calibration.
With
Miscellaneous
all other parameters are reset to the fac-
tory settings. This includes:
The range of the temperature limit values is reset to 81 °C
and 3 °C.
The baud rate is reset to
9600 bauds
.
Operating button locking is deactivated.
The temperature control unit is reset to
°C
.
The sound volume of the signal tone is reset to
Loud
.
The display brightness is reset to
Level 5
.
1.
Select the
Menu
Setup
Factory setting
menu item.
2.
Select the menu item for the parameter that you want to
reset.
3.
Press [OK] to confirm.
4.
Select the [Yes] menu item.
5.
Press [OK] to confirm.
The factory setting stored in the device is restored.
Restore factory calibration
Operation
LOOP
26