ID Function
Unit
Command
80 Continue programmer (after pause)
[–]
RMP_CONT
81 End programmer
[–]
RMP_STOP
4.7.8 Error messages
The error messages of the Ethernet interface are described in the following.
The string
ERR_X
or
ERR_XX
is output after an incorrect command.
Error
Description
ERR_2
Incorrect entry (for example, buffer overflow)
ERR_3
Wrong command
ERR_5
Syntax error in value
ERR_6
Impermissible value
ERR_8
Module or value not available
ERR_30 Programmer, all segments occupied
ERR_31
Not possible to specify a set point value. The "Setpoint offset"
function is active.
ERR_32
TiH
£
TiL
ERR_33
External sensor missing
ERR_34
Analog value not present
ERR_35
Automatically configured
ERR_36
Not possible to specify set point, programmer is running or
has been paused
ERR_37
Not possible to start programmer (analog set point value input
is ON)
ERR_38
The operator does not have the privileges to execute the
command. Another control section has exclusive operating
rights, which means that writing via this interface is not per-
mitted.
ERR_39
Operation is not permitted. Safe Mode is active.
ERR_40 Operation is not permitted. Safe Mode is switched off.
ERR_41
Operation is not permitted. Constant temperature equipment
is in error status.