n
Control station/PC
connected to the constant temperature equipment via Ethernet
interface, RS 232/485 interface (optional accessory), Profibus
interface (optional accessory) or EtherCAT interface (optional
accessory)
n
Analog interface (optional accessory)
n
Contact interface (optional accessory)
Allowing access to the device via the network
If you wish to have digital access to the device from the outside, this
must be set in the device software beforehand.
1.
Press the [Enter key] to open the menu.
2.
Select the menu items
Setup
Basic setup
Ethernet
PC
control
or
Web server
.
The options [off] and [on] appear on the display.
3.
Select the [on] option and press the Enter key to confirm.
A check mark is set. The entry has been accepted.
n
The full range of functions is available via the Master without restriction.
n
Operation via the Command is reduced by the following functions:
Entry of TMax
Selection of the heat transfer liquid
n
Operation via the Web server is reduced by the following functions:
Entry of TMax
Selection of the heat transfer liquid
Safety functions which require user presence at the device (e.g.
menus for filling and draining)
n
The control station is limited by the functionality of the interface and its
protocol (command set).
n
The analog interface and contact interface are restricted by their func-
tionality and protocol.
Applies equally to Master, Command, Web server and Control station via
n
Operator, once at maximum
The operator has all setting options available, both reading and
writing, if included in the range of functions of the operating unit.
n
Viewer, multiple times possible, only read privileges
All menus are accessible to the viewer, but no settings which change
the function of the device can be made. Exceptions are entries
which are necessary to log in as Operator.
The Master has Operator privileges in the set delivery condition.
If a user logs in via Web server or connects the Command operating unit, the
Web server or Command has Viewer privileges.
Allowing access to the device
Range of functions of the operating units
Operator and viewer
V6
Integral Process Thermostats and High-Temperature Thermostats
116 / 198