ECO GOLD
84
Appendix with settings
25/08/2011/
YACE0088
A.6
Limiting the mains current consumption
If your mains fusing is below 16 A, the current consumption can be reduced in steps from 16 A to 8 A.
The maximum heating power is reduced correspondingly. Here, take into consideration whether other
loads are connected to the same fused circuit or whether your ECO thermostat is the only load.
−
Access to the main menu level is obtained by pressing the
enter key
.
−
Selection and confirmation of
Æ
Setup
Æ
Basic setup
.
The adjacent menu window appears.
−
As well as
Curr.Consumpt.
the set value is displayed.
−
Select
Curr.Consumpt.
with
.
−
Change the value with or .
−
Single figures can be selected by pressing or .
−
Confirm the input with the enter key
.
−
By pressing (ESC) you are returned to the menu level without any change.
A.7
Entering the offset of the displayed temperature (calibration)
Deviations to the calibrated reference thermometers (e.g. LAUDA DigiCal) can be corrected internally
by the "Offset" function.
−
Access to the main menu level is obtained by pressing the
enter key
.
−
Selection and confirmation of
Æ
Setup
Æ
Calibration
.
The adjacent menu window appears.
−
Select
Calibration
with or and confirm with
.
The entry window appears. The value indicated on the refer-
ence thermometer must be entered as the value.
−
Change the value with or .
−
Single figures can be selected by pressing or .
−
By
pressing (+/-) the arithmetic sign can be changed.
−
You confirm the set value by pressing
.
−
By pressing (ESC) you are returned to the menu level without any change.
ESC
סּ
OK
Standby
Basic setup
Acoustic
signal
►
Display
►
Autostart on
►
Resolution
Curr.Consumpt. 16.0A
DLK
connected
ESC
סּ
OK
Standby
Curr.Consumpt.
Max:
16.00
Min: 8.00
16.00
ESC
סּ
OK
Standby
Calibration
Calibration
.
Factory calibration
ESC
סּ
OK
+/-
Temp. ref. meas. device
Max:28.16
Min:-22.16
25.16