ECO GOLD
25/08/2011/ YACE0088
Before putting the device into operation
21
6
Before putting the device into operation
Please note:
−
The device can be operated up to an ambient temperature of 40 °C.
−
A higher ambient temperature can have a negative effect on the cooling output of the thermostats used.
−
When putting the chiller into operation after a lengthy shut-down, up to 30 minutes may pass until the rated
refrigerating power is available depending on room temperature and device type.
6.1
Assembly and siting
Always comply with the following safety information:
Warning !
Falling / toppling equipment on sloping surfaces / table
edge
Crushing of hands and feet
•
Only site the device on flat surfaces, not near the
edge of the bench or table.
Affix the symbol "Hot surface".
The ECO thermostat is used as:
−
Immersion thermostat (optionally with cooling coil and/or pump connection set),
−
Heating thermostat (heating bath and circulation thermostat),
−
Cooling thermostat (cooling/heating bath and circulation thermostat).
Assembly as immersion thermostat
−
Push the screw clamp on the underside of the control
head into the guide rails.
−
Insert the thermostat with the screw clamp into the tem-
perature control vessel (
Ì
9) and screw the clamp
tightly to the bath edge with the knurled screw.
−
With plastic baths the tubular heating element must not
contact the bath wall.
−
Ensure that the ventilation opening at the back of the
device is free.
−
Keep a distance of at least 20 cm free on all sides of the
device.
Warning !
Control head drops into bath
Electric shock hazard
•
Make sure that the control head mounting is securely
joined to the bath.