![Larson Davis Spartan 821 Reference Manual Download Page 91](http://html1.mh-extra.com/html/larson-davis/spartan-821/spartan-821_reference-manual_3720611091.webp)
Spartan 821/721 Manual
IEC 61672-1 Section 9.3
B-2
TABLE B.1
821 with PRM821/377B02 Average Frequency Responses and Corrections,
Required by IEC 61672-1
Exact
Frequency
0° Free
Field
Response
0° Free-
Field
Corrections
1
Effect of
Wind
Screen
0° Free
Field
821 with
Wind
Screen
0° Free-
Field
Corrections
821 with
Wind
Screen¹
Expanded
uncertainty
of
Corrections
@ 95%
Hz
dB
dB
dB
dB
dB
dB
63
0.18
-0.18
-0.06
0.12
-0.12
0.25
79
0.18
-0.18
-0.06
0.12
-0.12
0.25
100
0.18
-0.18
-0.06
0.12
-0.12
0.25
126
0.18
-0.18
-0.06
0.12
-0.12
0.25
159
0.18
-0.18
-0.06
0.12
-0.12
0.25
200
0.18
-0.18
-0.06
0.12
-0.12
0.25
251
0.18
-0.18
-0.05
-0.01
0.01
0.25
316
0.18
-0.17
-0.02
0.02
-0.02
0.25
398
0.22
-0.22
-0.05
0.04
-0.04
0.25
501
0.24
-0.24
0.05
0.19
-0.19
0.25
631
0.28
-0.28
0.02
0.22
-0.22
0.25
794
0.11
-0.11
0.11
0.18
-0.18
0.25
1000
0.00
0.00
0.14
0.14
-0.14
0.25
1059
-0.08
0.08
0.19
0.11
-0.11
0.25
1122
-0.19
0.19
0.24
0.04
-0.04
0.25
1189
-0.20
0.20
0.17
-0.03
0.03
0.25
1259
-0.17
0.17
0.22
0.05
-0.05
0.25
1334
-0.34
0.34
0.32
-0.02
0.02
0.25
1413
-0.24
0.24
0.24
0.01
-0.01
0.25
1496
-0.30
0.30
0.34
0.03
-0.03
0.25
1585
-0.08
0.08
0.42
0.34
-0.34
0.25
1679
-0.15
0.15
0.44
0.28
-0.28
0.25
1778
0.14
-0.14
0.50
0.64
-0.64
0.25
1884
0.17
-0.17
0.49
0.65
-0.65
0.25