39
5.2 About This Meter
The About menu contains details about the meter hardware, software, and firmware, and
options; this includes regulatory compliance details, version, and issue dates.
Step 1.
On the meter, tap
Tools
About
to view the following information:
5.3 Troubleshooting the 721 or 821
The 721 or 821 System Utilities allow you to reboot the meter. This clears and resets the
meter for use. Note that unsaved measurement data on the meter will be saved. To reboot
the meter, do the following:
Step 1.
On the meter, select
Tools
System Utilities
.
Step 2.
Select
Reboot Meter
. The meter is unavailable for less than 10 seconds during
this process.
5.4 Troubleshooting File Storage
The 721 or 821 System Utilities allow you to troubleshoot issues with the memory card or
associated meter settings. Specifically, reformatting the microSD card and restoring meter
settings often resolves the issue. Note that unsaved measurement data on the meter is
also cleared.
CAUTION
Do not attempt to remove the microSD card while the meter is powered
on. This may result in data corruption or complete loss.
To reformat the microSD card and restore meter settings, do the following:
Step 1.
Power the meter off, and discharge accumulated electrostatic charge on your
body.
CAUTION
Do not open the microSD card access door while the meter is powered on.
This increases the risk of internal damage to the meter due to electrostatic
discharge (ESD).
Table 5.1 About This Meter: Details and Dates
Serial #
Certification Date
Hardware Version
Manufacture Date
International Standards Compliance
Firmware Version with Date and Install Date
Options Available for This Model
Options Currently Installed On This Meter
Regulatory section containing the certified Bluetooth® radio modules in the meter